Patient monitor User’s manual
14
Pediatric: 20 mmHg to 80 mmHg (normally 60mmHg);
Neonatal: 20 mmHg to 50 mmHg (normally 40mmHg).
The time after which cuff will deflate automatically:
180 sadult /pediatric, 90 sneonatal
Alarm range
Sys
0 mmHg to 300 mmHg, high/low limit can be
adjusted continuously.
Dia
0 mmHg to 300 mmHg, high/low limit can be
adjusted continuously.
Map
0 mmHg to 300 mmHg, high/low limit can be
adjusted continuously.
SunTech NIBP (option)
Way of measurement
Oscillometric. Diastolic values correspond to
Phase 5 Korotkoff sounds.
Range of measurement
Adult
SYS
40~260 mmHg
DIA
20~200 mmHg
MAP
26~220 mmHg
Child
SYS
40~160 mmHg
DIA
20~120 mmHg
MAP
26~133 mmHg
Neonate
SYS
40~130 mmHg
DIA
20~100 mmHg
MAP
26~110mmHg
Pressure messurement
Accuracy
± 3 mmHg (± 0.4 kPa) or 2 % of the reading above
200 mmHg for the ambient temperature range of
50to 104(10 to 40), the humidity range
of 15 % to 90 %
Pressure Transducer
Accuracy
±3mmHg between 0 mmHg and 300 mmHg for
operating conditions between 0 and 50.
Unit
mmHg, kPa
Pulse rate range
30 ~ 220 bpm
Pulse Rate Accuracy
2% or 3 bpm, whichever is greater
Inflation time for cuff
<75s