Patient monitor User’s manual
18
SpOC
Measurement range
0 ml/dl to 35 ml/dl
A.6.5 Temp
Standard
EN 12470-4
Measurement way
Thermal resistance way
Measurement range
0.0 to 50.032 to 122
Accuracy
±0.1
or ±1
(exclusive of probe)
Resolution
0.1
or 1
Unit
or
Data update period
Every 1 to 2seconds
Minimum accurate
measuring time
Surface: <100s;
Coelom: <80s.
Self check
About every 5min to 10 min
Temp sensor
resistance
2252Ω
25
Alarm range
0.0
to 50.0
, high/low limit can be adjusted
continuously.
A.6.6 IBP
Standard
EN 60601-2-34 / IEC 60601-2-34
Measurement way
Directly invasive pressure measurement
Sensitivity of
transducer
5uV/V/ mmHg, ±2%
Impedance of
transducer
300Ω to 3000Ω
Static pressure
measurement range
-50 mmHg to +350 mmHg