Patient monitor User’s manual
22
Warm-up time
Concentrations are reported and the automatic agent
identification is running within 10s. Full accuracy
within 10s.
Rise time(at 10
L/min)
90 ms
Total system
response time
< 1 s
CO
2
Measurement
Range
0 ~ 15%
CO2 accuracy
±0.3%
ABS
+4%
REL
Sidestream CO2 module (ISA)
Sampling flow rate
50ml/min ± 10ml/min
Compensations
Automatic compensation for pressure and
temperature.
Manual compensation for broadening effects on CO
2
.
Calibration
No span calibration is required for the IR bench. An
automatic zero reference calibration is performed at
startup and then every 24 hours.
Warm-up time
< 10 s (Concentrations reported and full accuracy)
Typical rise time
at 50 ml/min sample
flow
200 ms
Total system
response time
< 3 s (with 2 m sampling line)
CO
2
Measurement
Range
0-15%
CO2 accuracy
±0.3 kPa + 4% of reading
A.6.8 AG
Mainstream AG module (IRMA)
Measurement way
Infrared spectrum
Measurement mode
Mainstream
Measurement
parameters
CO
2
N
2
OagentISOENFSEVHALDES