Patient monitor User’s manual
23
Resolution
1%
Calibration
Zeroing recommended when changing Airway adapter. No
span calibration required for the IR bench.
Warm up time
Concentrations are reported and the automatic agent
identification is running within 10s. Full accuracy within
20s for IRMA AX+.
Rise time
(When the flowing
speed is 10 L/min)
CO
2
90 ms
N
2
O 300 ms
ISO, ENF, SEV,HAL,DES 300 ms
Total system
Response Time
1s
Primary agent
threshold
(IRMA AX+)
0.15%. When an agent is identified, concentrations will
be reported even below 0.15% as long as apnea is not
detected.
(When the concentration has passed the threshold, the
concentration will be reported even below 0.3%.)
Secondary agent
threshold
(IRMA AX+)
0.2%+10% of total agent concentration
Agent identification
time(IRMA AX+)
< 20s (typically < 10s)
Measurement range and accuracy of gas
Gas
Range (%)
Accuracy
1)
IRMA AX+
CO
2
0 to 15
±
0.3%
ABS
+4%
REL
N
2
O
0 to 100
±2%
ABS
+5%
REL
O
2
0 to 100
±2%
ABS
+2%
REL
ISO
ENF
HAL
0 to 8
±
0.2%
ABS
+10%
REL
SEV
0 to 10
±0.2%
ABS
+10%
REL
DES
0 to 22
±0.2%
ABS
+10%
REL
awRR measurement
range
0 rpm to 150 rpm
awRR measurement
accuracy
±1 rpm