Patient monitor Users manual
56
Physiological alarm:
Alarm messages
Cause
Level
CSI High
CSI measuring value is above the
high alarm limit
Medium
User-Selectable
CSI Low
CSI measuring value is below the
low alarm limit
Technical alarm:
Alarm messages
Cause
Level
Artefact
The interference is too large.
No alarm
Electrode alarm
There are some problems with
electrode, please check the
electrode.
SQI low
The signal quality is low.
Impedance high
Sensor impedance is too high.
Replace CSM Battery
Batteries are empty or broken,
please replace the battery.
Check Cable
The cable is not connected, please
check the connection.
Battery Low
The battery voltage is too low.
BIS
Physiological alarm:
Alarm messages
Cause
Level
BIS high
The BIS measurement is higher than
the set upper alarm limit
Medium
User-Selectable
BIS low
The BIS measurement is lower than
the set lower alarm limit.
Technical alarm:
Alarm messages
Cause
Level
BIS Sensor off
The CSI measurement is higher than
the set upper alarm limit.
Low
BIS Simulator
Connected
The Demo prob is not connected well
Use too Long Time
The prob use more than 24 hours
BIS Bad SQI
SQI signal value less of 15.
No Alarm
BIS Poor SQI
SQI signal value less of 50.