Patient monitor User’s manual
Chapter 6-5
2. If Neither Patient Data Set is Correct
SelectNew Patientin the asking window, if you are sure that none of the
information is correct. This discharges all patients, erases all data in both the
monitor and EMS, resets all settings to the default setting, and lets you admit a
new patient.
3. If Both Patient Data Sets are Correct
Select Same Patient in the asking window, if the patient information is
different, but you are sure it is the same patient. This merges the data stored in
EMS and monitor. EMS continues monitoring patient and the patient data, trend
data and measurement settings will be uploaded from the EMS to the monitor.