Patient monitor User’s manual
Chapter 7-14
7.7.3 Setting Arrhythmia Alarm
Select
ARR Analysis
ARR Alm Setup
in the ECG setting menu,
in the popped up menu, you can set the alarm status for each Arrhythmia,
including alarm switch, alarm level and alarm print switch. You can select
ALL ARR Alm-Resume Default
,
All ARR Alm-Alm Switch
,
ALL
ARR Alm-Alm LevelandAll ARR Alm-Alm Printto set the alarm status
for all arrhythmia.
Arrhythmia can be analyzed by the monitor is shown in the following table:
Arrhythmia Type
Applicable Patient
Type
Alarm Level
ASYSTOLE
All
Default High
VENT FIB/TACH
Non-paced
Default High
PAC
Non-paced
Default medium,User selectable
RUN PVCS
Non-paced
Default medium,User selectable
COUPLET
Non-paced
Default low,User selectable
BIGEMINY
Non-paced
Default low,User selectable
TRIGEMINY
Non-paced
Default low,User selectable
R ON T
Non-paced
Default low,User selectable
TACHY
All
Default medium,User selectable
BRADY
All
Default medium,User selectable
MISSED BEATS
Non-paced
Default low,User selectable
PNP
Paced
Default medium,User selectable
PNC
Paced
Default medium,User selectable
NOISE
All
Default low,User selectable
VPB
Non-paced
Default low,User selectable
ST HIGH
All
Default low,User selectable
ST LOW
All
Default low,User selectable
VTACH
Non-paced
Default High
PVCS HIGH
Non-paced
Default low,User selectable
7.7.4 Arrhythmia Relearning