Chapter 6 Analyzing Data in an RQ Study
Adjusting the Baseline and Threshold
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Relative Quantification Getting Started Guide 67
Notes
6
Threshold Set Correctly
The threshold is set in the exponential phase of the
amplification curve.
Threshold settings above or below the optimum
increase the standard error of the replicate groups.
Threshold Set Too Low
The threshold is set below the exponential phase of
the amplification curve. The standard error is
significantly higher than that for a plot where the
threshold is set correctly. Drag the threshold bar up
into the exponential phase of the curve.
Threshold Set Too High
The threshold is set above the exponential phase of
the amplification curve. The standard error is
significantly higher than that for a plot where the
threshold is set correctly. Drag the threshold bar down
into the exponential phase of the curve.

Table of Contents

Other manuals for Applied Biosystems 7500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7500 and is the answer not in the manual?

Applied Biosystems 7500 Specifications

General IconGeneral
Detection MethodFluorescence
Sample Capacity96-well plate
Thermal CyclerYes
Temperature Uniformity±0.5°C
TypeReal-Time PCR System
Thermal Accuracy±0.25°C
Detection Channels5
Weight45 kg
Power Requirements100-240 V, 50/60 Hz
Dynamic Range9 logs
Reaction Volume Range10-100 µL
Temperature Range4°C to 100°C
Thermal Cycler Block96-well
SoftwareSDS Software
DetectionMultiplex (up to 5 colors)

Related product manuals