34
This parameter defaults to one year after the calibration date but may be overwritten to any
advanced date desired. Within the Calibration Due Date parameter are three separate fields,
month, day, and year. Use the > key to select the field within the date you want to edit then
use the + key to change the numeric value. Once you change the value, it is automatically
stored.
Alert (Alert Date)
The Alert date is like an alarm clock that will warn you in advance of the actual calibration due
date. After a calibration is performed, the Alert Date is automatically set 11 months after the
calibration date. For example, if the calibration is performed on 12/15/2002 the Alert Date will
automatically be set to 11/15/2003.
This parameter defaults to 11 months after the calibration date but may be overwritten to any
advanced date desired. Within the Alert Date parameter are three separate fields, month, day,
and year. Use the > to select the field within the date you want to edit. Use the + key to
change the numeric value. Once you change the value, it is automatically stored.
Date
Scroll the cursor to the Date parameter using the directional soft keys.
Within the Set Date parameter are four fields, date format (mdy / dmy), month, day, and year.
Use the > key to select the field within the date you want to edit. Using the + key to change
the numeric value. Once you change the value, it is automatically stored.
Time
Scroll the cursor to the Time parameter using the directional soft keys.
Within the Set Time parameter are three fields, hours, minutes, and Military or Standard time
setting. Use the > key to select the field within the time parameter you want to edit. Use the +
key to change the numeric value. Once you change the value, it is automatically stored.
The Calibration Alert Warning screen appears as follows:
At the Calibration Alert Warning screen, there are two options, “Show screen again? Yes or
No”. “Show screen again?” is asking if you would like to continue seeing the Calibration Alert
Cal. Alert
Model 36XX is due
for calibration on
10/16/04.
Show screen again?