65
Calibration of DC Current 7.50mA range
Calibration of DC Current 3.50mA range
Calibration of DC Current 350µA range
Calibration of IR 999.9M and 9999M Ranges
Calibration
+
Enter standard
Current reading
Current= XXXXV
_
<RESET> to Exit
Enter
Calibration
Connect the 100K
load in series with
the 7.5mADC standard
current meter.
<TEST> to Start
<RESET> to Exit
Calibration
+
Enter standard
Current reading
Current= XXXXV
_
<RESET> to Exit
Enter
Calibration
Connect the 100K
load in series with
the 3.5mADC standard
current meter.
<TEST> to Start
<RESET> to Exit
Calibration
+
Enter standard
Current reading
Current= XXXXV
_
<RESET> to Exit
Enter
Calibration
Connect the 1M
load
in series with the
350uADC standard
current meter.
<TEST> to Start
<RESET> to Exit
Calibration
Connect the standard
50M
load from H.V.
to Return.
<TEST> to Start
<RESET> to Exit
Calibration
Automatic
calibration
processing...