15Hz < f < 100KHz : ±5% of reading
(0.01V -10.00V )
DISPLAY FIXED RANGE (<6% OF RANGE)
Touch Voltage Display (RMS) (AC
Only)
0.0mV ~ 7.8mV, frequency
15Hz - 1MHz
15Hz < f < 30Hz : add ±(2% of reading
+ 0.2% of range) to base accuracy
30Hz < f < 100KHz: add ±(2% of
reading + 0.2% of range) to base
accuracy
100KHz < f < 1MHZ : add ±(2% of
reading + 0.5% of range) to base
accuracy
0.0mV ~ 31.5mV, frequency
15Hz - 1MHz
15Hz < f < 30Hz : add ±(2% of reading
+ 0.2% of range) to base accuracy
30Hz < f < 100KHz: add ±(2% of
reading + 0.2% of range) to base
accuracy
100KHz < f < 1MHZ : add ±(2% of
reading + 0.5% of range) to base
accuracy
0.0mV ~ 126.0mV, frequency
15Hz - 1MHz
15Hz < f < 30Hz : add ±(2% of reading
+ 0.2% of range) to base accuracy
30Hz < f < 100KHz: add ±(2% of
reading + 0.2% of range) to base
accuracy
100KHz < f < 1MHZ : add ±(2% of
reading + 0.5% of range) to base
accuracy
0.0mV ~ 504mV, frequency
15Hz - 1MHz