207
You may now store the new calibration number by pressing the ENTER key or escape
by pressing the EXIT key or the RESET button.
Calibration of IR 999.99M Range
The load/meter screen and prompt screens will contain the following text for this
calibration point:
Upper Right Screen
Load/Meter
Lower Right Screen
Prompt
Lower Right Screen
Prompt (Cal. Active)
Connect the standard
50M load from H.V. to
Return.
When the standard load is
connected, press TEST to
start the calibration
process.
Automatic Calibration
Processing.
Press the TEST button to activate the calibration. After the calibration is activated, the
lower right screen will change to the text indicated in the table above (Cal. Active).
This calibration is automatic and does not require data entry.
Calibration of IR 9999.9M Range
The load/meter screen and prompt screens will contain the following text for this
calibration point:
Upper Right Screen
Load/Meter
Lower Right Screen
Prompt
Lower Right Screen
Prompt (Cal. Active)
Connect the standard
50M load from H.V. to
Return.
When the standard load is
connected, press TEST to
start the calibration
process.
Automatic Calibration
Processing.
Press the TEST button to activate the calibration. After the calibration is activated, the
lower right screen will change to the text indicated in the table above (Cal. Active).
This calibration is automatic and does not require data entry.
Calibration of IR 50000M Range
The load/meter screen and prompt screens will contain the following text for this
calibration point:
Upper Right Screen
Load/Meter
Lower Right Screen
Prompt
Lower Right Screen
Prompt (Cal. Active)
Connect the standard
500M load from H.V. to
Return.
When the standard load is
connected, press TEST to
start the calibration
process.
Automatic Calibration
Processing.
Press the TEST button to activate the calibration. After the calibration is activated, the
lower right screen will change to the text indicated in the table above (Cal. Active).
This calibration is automatic and does not require data entry.