62
4.4.2. Default Test Parameters
In each of the Test Parameter Setting screens there is a “Defaults” soft key. When
this key is pressed the test will load with a set of predetermined Default Parameters.
The following table is a listing of the Default Parameters for each of the different types
of tests that are available in the OMNIA II. The Default Parameters are as follows:
Default Parameters
TEST TYPE
PARAMETER
VALUE
ACW
Voltage
1240VAC
HI-Limit T
10.00mA
LO-Limit T
0.000mA
Ramp Up
0.1s
Dwell Time
1.0s
Ramp Down
0.0s
Arc Sense
5
HI-Limit R
10.00mA
LO-Limit R
Offset
0.000mA
0.000mA
Scanner
OOOOOOOOOOOOOOOO
Frequency
60Hz
Arc Detect
OFF
Continuity
OFF