196
9.1.3. AC Hipot Verification
From the FailCHEK screen, press the “AC Hipot” soft key. A pop-up message will
now be displayed. The message will appear as follows:
Follow the instructions given in the pop-up message and then press TEST to begin
the verification process. During the Verification process the metering screen for the
related process will be displayed, refer to section 5.4. Perform Tests Metering for
more information.
At the end of the Verification process, a pop-up message will appear indicating the
outcome of the process. If the instrument passes the Verification (test failure,
indicating the fail detectors are working properly) the RESET button will illuminate, the
alarm will sound and the following pop-up message will appear over the metering
screen: