24
DISPLAY (CONT.)
Range 4
400uA ~ 2100uA, frequency
DC, 15Hz - 1MHz
Resolution
1uA
Accuracy
DC , 15Hz < f <100KHz: ±(2%
of reading + 3counts)
100KHz < f < 1MHZ : ±5% of
reading (10uA - 8500uA)
Range 5
1800uA ~ 8500uA, frequency
DC, 15Hz - 1MHz
Resolution
1uA
Accuracy
DC , 15Hz < f <100KHz: ±(2%
of reading + 3counts)
100KHz < f < 1MHZ : ±5% of
reading (10uA - 8500uA)
Range 6
8.00mA ~ 10.00mA, frequency
DC, 15Hz 100kHz
Resolution
0.01mA
Accuracy
DC, 15Hz < f < 100KHz : ±5%
of reading (0.01mA -10.00mA )
Touch Current Display (Peak)
Range 1
0.0uA ~ 32.0uA, frequency DC -
1MHz
Resolution
0.1uA
Accuracy
DC : ±(2% of reading + 2uA)
15Hz < f < 1MHZ : ±10% of
reading + 2uA
Range 2
28.0uA ~ 130.0uA, frequency
DC - 1MHz
Resolution
0.1uA
Accuracy
DC : ±(2% of reading + 2uA)
15Hz < f < 1MHZ : ±10% of
reading + 2uA
Range 3
120.0uA ~ 550.0uA, frequency
DC - 1MHz
Resolution
0.1uA
Accuracy
DC : ±(2% of reading + 2uA)
15Hz < f < 1MHZ : ±10% of
reading + 2uA

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Associated Research OMNIA II 8207 and is the answer not in the manual?

Associated Research OMNIA II 8207 Specifications

General IconGeneral
BrandAssociated Research
ModelOMNIA II 8207
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals