82
LINE
CONFIG.
NEUTRAL
REVERSE
GROUND
Relay/
Soft
key
Fault
Soft key
Relay
Fault
Relay/
Soft key
Fault
1
Open
Yes
OFF
A
No
Open
Yes
2
Open
Yes
ON
B
Yes
Open
Yes
3
Open
Yes
OFF
A
No
Closed
No
4
Open
Yes
ON
B
Yes
Closed
No
5
Closed
No
OFF
A
No
Open
Yes
6
Closed
No
ON
B
Yes
Open
Yes
7
Closed
No
OFF
A
No
Closed
No
8
Closed
No
ON
B
Yes
Closed
No
Probe Configuration
The Probe configuration is controlled by the relays SH and SL shown in Figure 2 and
are set by the “Probe” soft key. These two relays configure the current measuring
device into three possible positions. Repeatedly pressing the Probe soft key allows
the selection of the three different states. The three positions are described in the
table below.
MEASURING
DEVICE POSITION
SH
SL
DISPLAY
INDICATION
AGENCY TEST
APPLICATION
Ground to Line
A
A
Ground to Line
Protective Conductor
Current
Probe HI to Line
B
A
Probe-HI to Line
Touch Current
Probe-HI to Probe-LO
B
B
Probe-HI to Probe-
LO
External Applied Part or
Point to point Leakage
A diagram of the probe configurations is shown in the image below.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Associated Research OMNIA II 8207 and is the answer not in the manual?

Associated Research OMNIA II 8207 Specifications

General IconGeneral
BrandAssociated Research
ModelOMNIA II 8207
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals