76 E1070 Service Manual
3 Using the Menus
12. Toggle between the choices by using the TARE or UNITS key. Press ENTER
when your choice is displayed…
bASiS is displayed.
13. Press the UNITS key…
CAL is displayed. Use this to calibrate the analog output.
14. Press the ENTER key…
ZEro is displayed. Use this to set the analog output zero point.
15. Press the PRINT key…
AdJuSt is displayed. Use this to adjust the zero output up or down by
pressing the SELECT (up) or PRINT (down) key. Each key press
changes the output by 0.25% of available range for the output.
When you are done, press the ENTER key and go to step 19.
If you need to change the zero output by a custom amount, press the
ENTER key and go to step 16.
16. To change the zero output by a custom amount, from the AdJuSt display,
press the UNITS key…
CUSt. is displayed.
17. Press the PRINT key…
A numeric entry screen appears.
18. Enter a custom amount, in percentage of available range for the output, and
press the ENTER key…
CUSt. is displayed.
19. Press the SELECT key…
ZEro is displayed.
20. Press the UNITS key…
SPAn is displayed. Use this to set the analog output span point.
21. Repeat steps 15-18 to adjust span.
22. Repeatedly press the SELECT key until…
A-Out is displayed.
CNTR (Pulse counter ) Optional circuitry required
Follow these steps to configure the pulse counter.
1. With A-Out displayed, press the UNITS key…
Cntr is displayed. This stands for the pulse counter option. Use this item
to enable and enter a factor for converting pulses into your unit of
measure.
2. Press the PRINT key…
on or oFF is displayed.