ENGINEERING DOCUMENTATION 31
_____________________________________________________________________________________________
12.9 Percentage weighing function (PErC)
This function allows displaying
weighing result in percents.
A measurement is performed in two
phases:
- first phase – weighing a reference
sample (100%),
- second phase – measuring specific
sample as a percentage of
the reference sample.
Weighing result is displayed in
different format, depending on the
weight value of reference sample. For
weight values of reference sample
0÷3,5% of weighing range result is
displayed in format 100, for range
3,5÷35% - in format 100.0, and
above 35% - in format 100.00.
„%” sign is replaced with „
”
indicator.
The function has the following
options:
- PErC-0 – disable the function,
- PErC-1– set current scale indication
as 100% and activate percentage
weighing,
- PErC-2 – continue percentage
weighing after exiting to normal
weighing.
Note:
During percentage weighing
→
T
←
key has its normal function.
95.00
100%
5%
_
_
_
_