EasyManua.ls Logo

BIO RAD Tetrad 2 - Editing a Step; Inserting a Step

BIO RAD Tetrad 2
148 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Managing and Editing Programs
Tech Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com 6-7
Inserting a Step
Now we will insert an annealing temperature step.
• Highlight step 2.
Recall that new steps are inserted AFTER the highlighted step.
• Select <Temp> from the advanced programming window.
The Temperature Step window appears. Select an incubation Temperature of
60°C, a hold Time of 3 minutes, and the Beep on Target option. Enter param-
eters as described in Chapter 5.
Your program should appear as follows.
Temperature Control Mode: Calculated
Lid Control Mode: Constant at 100°C
1. Incubate at 94°C for 00:01:00
Beep on Target
2. Incubate at 92°C for 00:00:30
Beep on Target
3. Incubate 60°C for 00:03:00
Beep on Target
4. Cycle to step 2 for 29 more times.
5. Incubate at 10°C forever
End
Editing a Step
Step 2 of the program ADV#1 includes a 30 second incubation. We wish to change
that incubation to 25 seconds.
• Select Step 2 by double-clicking.
You are presented with the Temperature Step window. The Temperature field
shows 92°C, and the Time field shows 30 seconds.
• Select the Time:Sec field.
• Change the time from 30 to 25.
• Select <OK>.

Table of Contents