Handheld monitor user’s manual
Chapter 5-6
Caution: During the monitoring of HR and PR,
displaying of HR has priority. That is PR will be
displayed only when there isn’t HR monitoring.
5.6 SpO
2
Alarm Setup
5.6.1 Switching On/Off SpO
2
Alarm
1. Select
Menu
Alarm Setup
.
2. Set theAlarmof SpO
2
toOffto shut off SpO
2
alarm.
When the alarm of SpO
2
is off, there is a sign of in
the SpO
2
parameter display area.
5.6.2 Setting Alarm Level
1. Select
Menu
Alarm Setup
.
2. Set theAlarmof SpO
2
toMedorHigh.
5.6.3 Adjusting the Alarm Limit
1. SelectMenu】→【Alarm Setup.
2. AdjustHigh:If the SpO
2
measurement is higher than
the high alarm limit, the “SpO2 Too High” alarm will be
triggered.