Handheld monitor user’s manual
Chapter 9-4
Test the patient leakage current Mains voltage on applied
part: According IEC 60601-1:1988 + A1:1991 + A2:1995:
Limit: type CF: for a.c.: 50uA.
Warning: No use-serviceable parts inside, before
servicing to authorized representative or
manufacturer.
9.3 Cleaning the Monitor
1. Common detergent and non-corrosive disinfectant used in
hospital can be applied to clean monitor, however you
must be aware that many kinds of detergents must be
diluted prior to utilization, and please use it according to
the instruction of detergent manufacturer.
2. Avoid the use of alcohols, amino or acetonyl detergent.
3. The enclosure and screen of monitor shall be free of dust,
and they can be wiped with lint-free soft cloth or sponge
soaked in detergent. While cleaning, be careful and do not
spill liquid onto the instrument and keep any liquid out of
it. When wiping the side panel of monitor, you must be