Chapter 9-3
The data should be recorded in an equipment log. If the
device is not functioning properly or fails any of the
following tests, the device has to be repaired.
Inspect the equipment and accessories for mechanical and
functional damage.
Inspect the safety relevant labels for legibility.
Verify that the device functions properly as described in
the instructions for use.
Test the earth leakage current according IEC
60601-1:1988 + A1:1991 + A2:1995: Limit: NC 500μA,
SFC: 1000μA.
Test the enclosure leakage current according to IEC
60601-1:1988 + A1:1991 + A2:1995: Limit: NC 100μA,
SFC: 500μA.
Test the patient leakage current (normal operation)
according IEC 60601-1:1988 + A1:1991 + A2:1995:
Limit: type CF: for a.c.: 10μA, for d.c.: 10μA.
Test the patient leakage current under single fault
condition according IEC 60601-1:1988 + A1:1991 +
A2:1995:
Limit: type CF: for a.c.: 50μA, for d.c.: 50μA.