Handheld monitor user’s manual
Chapter 6-9
6.7 Hum filter Setup
1. Select
【
Menu
】→【
System
】→【
Maintenance
】
→enter the required password.
2. Select【Factory Setup】→【ECG Maintenance】
→【
Hum
】
, Set its value to the local main power frequency.
6.8 HR Alarm Setup
6.8.1 Alarm Switch Setup
Select
【
Menu
】→【
Alarm Setup
】
, Set
【
Alarm
】
of HR
to【Off】to shut off HR alarm. There is a sign of shown
in the left corner of HR display area. Or the HR alarm is on.
6.8.2 Alarm Level Setup
Select【Menu】→【Alarm Setup】, Set【Alarm】of HR
to
【
Low
】
,
【
Med
】
or
【
High
】
.
6.8.3 Alarm Limit Setup
Select
【
Menu
】
→
【
Alarm Setup
】
, set the high/low
alarm limit of HR to a value as required. The high alarm
limit should greater than the low one.