Handheld monitor user’s manual
Chapter 5-4
5.3 Monitoring Procedure
1. Selecting SpO
2
Sensor
Depending on the patient category, weight and
application site, you can select the SpO
2
sensor as required.
2. Connecting SpO
2
Sensor
Plug the SpO
2
sensor cable into the SpO
2
connector on
the measurement module.
3. Applying SpO
2
Sensor
Clean the application site, such as colored nail polish,
and apply the sensor to the patient.
Warning:
Do not use the SpO
2
sensor on a limb where the
NIBP cuff is applied. This may result in inaccurate
SpO
2
reading due to blocked blood flow during cuff
inflation.
Do not conduct SpO
2
measurement on the finger
smeared with nail polish, otherwise unreliable
measurement results might be produced.
When using finger sensor, make sure the nail faces
to the light window.