Handheld monitor user’s manual
11
sensitivity
Scan speed
12.5mm/s25 mm/s
Protection
Electric isolation voltage is 4000V at
50Hz/60Hz., defibrillator-proof.
Defibrillator-proof
Recovering time for defibrillator-proof ≤5s
Recovering time of
electrode polarization
after defibrillation
ECG waveform will recover to the baseline
within 10 seconds.
Pacemaker pulse
rejection amplitudes
and widths
Without
Electrosurgery
protection
( electrotome
protection )
Cut power :300W
Coagulation power :100W
Change of heart rate : ≤10
Recover time : ≤10s
Comply with the requirement of clause
4.2.9.14 in ANSI/AAMI EC 13:2002
Tall T-Wave rejection
capability
Minimum recommended 1.2 mV T-Wave
amplitude. Comply with requirement of
clause 4.1.2 c in ANSI/AAMI EC 13:2002.)
Heart rate averaging
50 bpm, once every two beats;