O-9a Rev F
operatIon
O-9a
No number t o the left of the test indicates tests which cannot be altered.
Example 3 CKMB indicates that CKMB is always selected, and cannot be
deselected.
• Pressthe key.
• Iftheinformationiscorrect,pressthe key to continue.
• Iftheinformationisincorrect,selectcorrect teSt using the
keys
and press the key to return to the lists of tests.
Important InStructIonS For all teStS:
a. Gently insert the test device into the Triage
®
MeterPro
®
Meter until you feel
the device catch on the pin. You will hear an audible ‘click’ that tells you
the test device has been inserted properly.
b. Press the key to start the test.
c. The Triage
®
MeterPro
®
Meter pulls in the test device and scans it.
d. When the test is complete, the meter will beep, eject the device and
display the results on the meter’s screen.
e. Press the key to make a printed copy of the results.
Important InStructIonS For patIent teStS:
a. If the test device’s Internal Quality Control (QC) results are unacceptable,
the results for the affected analyte(s) will not appear on the screen (instead
will be flagged with a ! symbol).
b. If any of the QC Sample results for the lot being tested were out of range,
the results for that particular analyte will not appear on the screen (instead
will be flagged with a # symbol).
Note: If no key
is pressed on
the Select Tests
screen, the meter
waits 30 seconds then
proceeds using only
the selected default
tests. Similarly, if no
key is pressed on the
Confirmation Screen,
the test proceeds after
waiting 30 seconds.
Screen displaying patients results
below the cut off levels and
accept or reject instructions.
Screen displaying patients results
above the cut off levels and TNI
results in the indeterminate range.
Afterthetest
device has been
inserted, do not
push the device in
further or attempt to
pull it out. The device
may be ejected by
returning to the Main
Menu and pressing
EXIT.