O-21 Rev F
operatIon
O-21
uSer Id
To retrieve QC results from tests run on the Triage MeterPro Meter starting with the user you
select:
1. Press the key to power The Triage MeterPro Meter on.
2. From the maIn menu select recall reSultS using the
keys.
3. Press the key.
4. Select Qc reSultS using the
keys.
5. Press the key.
6. Select the appropriate menu, Qc Sample reSultS or Qc devIce reSultS
using the
keys.
7. Press the key.
8. If Qc Sample reSultS, select panel type using the keys.
Use the keys to select the desired panel.
9. Select USER ID using the keys.
10. Press the key.
11. Type in the user ID. To correct, press the key to clear the entire
ID, or select the incorrect character by using the keys. Then
type in the correct ID.
12. Press the key.
13. All QC Sample or QC Device results from tests run by the requested user, for the
selected panel if QC Sample, are displayed.
14. Use the keys to view the entire record.
15. Press the key to make a printed copy.
16. Press the key to upload the selected results to LIS.
maIn menu
recall reSultS
Qc reSultS
Qc Sample reSultS
Qc devIce reSultS