O-20Rev F
operatIon
O-20
devIce lot numBer (l/n)
To retrieve QC Sample results based on the lot number of the test devices used:
The lot number of the test device is found on
the device and on the outer box
1. Press the key to power The Triage MeterPro Meter on.
2. From the maIn menu select recall reSultS using the
keys.
3. Press the key.
4. Select Qc reSultS using the
keys.
5. Press the key.
6. Select Qc Sample reSultS using the
keys.
7. Press the key.
8. Select panel type using the
keys.
9. Use the keys to select the desired panel.
10. Select devIce l/n using the keys.
11. Type in the device lot number from the side of the box of test devices. To correct,
press the key to clear the entire number. Then type in the correct number.
12. Press the key.
13. All QC results that have been run on the requested lot number for the selected
panel are displayed.
14. Use the keys to view the entire record.
15. Press the key to make a printed copy.
16. Press the key to upload the selected results to LIS.
Note: When
printing
large quantities of
patient results, it is
recommended to
first insert a full roll
of paper.
maIn menu
recall reSultS
Qc reSultS
Qc Sample reSultS