34 | Chapter 4: Instrument Qualification
Stock Solution Formulation
The stock solution for Liquid Tests 1 and 2 may be formulated from the chemicals
listed below, or by diluting a dye solution available from BioTek. See Procedure A
or Procedure B outlined below and on the following page for details.
Procedure A
Required Materials:
BioTek QC Check Solution No. 1 (PN 7120779, 25 mL; or 7120782,
125 mL)
Deionized water
5-mL Class A volumetric pipette
100-mL volumetric flask
Preparation of Stock Solution:
1.
Pipette a 5-mL aliquot of BioTek QC Check Solution No. 1 into a 100-mL
volumetric flask.
2. Add 95 mL of DI water; cap and shake well. The solution should measure
approximately 2.000 OD when using 200 µL in a flat-bottom microwell.
This should create a solution with an absorbance of about 2.000 OD when
using 200 µL in a flat-bottom microwell. The OD value will be proportional
to the volume in the well and the amount of QC Check Solution No. 1 used.
You can use a larger or smaller well volume, or add more Check Solution or
water to adjust the stock solution.
Too small a well volume may result in increased pipetting-related
errors.
Procedure B
Required Materials:
Deionized water
FD&C Yellow No. 5 dye powder (typically 90% pure)
Tween 20 (polyoxyethylene (20) sorbitan monolaurate) or
BioTek wetting
agent, PN 7773002
Precision balance with readability of 0.001 g
1-liter volumetric flask
Weigh boat
BioTek
®
Instruments, Inc.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Biotek Epoch and is the answer not in the manual?

Biotek Epoch Specifications

General IconGeneral
TypeMicroplate Spectrophotometer
Wavelength Range200 - 999 nm
Light SourceXenon flash lamp
Power Requirements100 - 240 VAC, 50/60 Hz
Wavelength Repeatability+/- 0.2 nm
Absorbance Range0-4.0 OD
Plate Types6- to 384-well microplates
ShakingLinear, orbital
Detection ModesAbsorbance
SoftwareGen5
Bandwidth5 nm
ConnectivityUSB

Related product manuals