64 | Appendix B: Error Codes
Code Description and Probable Causes
2Dnn
Assay error.
This error indicates that something about the assay that is
being attempted is wrong. Refer to the list of assay errors
that follows. The most probable cause of these errors is
something in the Gen5 protocol or instrument selection.
2A0x
Plate jam error.
Motor “x” (0 = X, 1 = Y, 2 = F, 3 = M) has hit something and
lost steps. Usually this is because the carrier was pushed or
restricted on its way into the instrument. Restart the read.
Assay Errors—For 2Dnn Error Code
Error Code Description
00 SAMPLE_START_LATE—missed start of well mode sample
01 INVALID_READSETS—number of readsets defined invalid
05 INVALID_MODE—invalid mode selection received
06 ASSAY_DEF_NOT_SET—validated assay not defined
07 INVALID_ROW_COL—invalid number of rows and columns
0B INVALID_EVENT_COUNT—total events defined invalid
0C INVALID_EVENT_TYPE—invalid event type received
0E PLATE_START_LATE—missed start of plate mode event
12 INVALID_WELL_DATAPTS—number of samples out of range
13 INVALID_WELL_INTERVAL—well mode kinetic interval out of
range
14 INVALID_PLATE_DATAPTS—specified # of plate mode reads
out of range
15 INVALID_PLATE_INTRVAL—plate mode kinetic interval out of
range
16 READ_START_LATE—missed start of read
17 INTERVAL_START_LATE—missed start of kinetic interval setup
43 INVALID_DIRECTION—initial row or column direction is invalid
44 DEBUG_MODE—unit is in debug mode; self-test may not have
been run