General Errors | 57
General Errors
A lowercase letter in the code indicates a variable numeral.
Code Description and Probable Causes
*0210
*02x1
The x-axis home opto sensor failed to change as
expected.
A y-axis opto sensor failed to change as expected.
x = 1 for the home sensor.
x = 2 for the test sensor.
Possible causes:
Dirty axis rail where the bearings are worn and cause
too much friction.
Defective or broken optical sensors or connection.
Defective motor, motor controller, or connection.
Broken or untensioned drive belt.
Obstruction in read chamber.
Check that the shipping screw has been removed.
Note: In cases where a sensor is not functioning, the motor
may drive the axis to its mechanical stop and generate
substantial noise but will not cause permanent damage.
*0212
The order sorting filter wheel’s home opto sensor failed
to change as expected.
Probable causes:
Filter wheel is loose.
Filter wheel gearing is worn or dirty.
Defective or broken optical sensor or connection.
Defective motor, motor controller PCB, or cable.
0233
Monochromator not able to locate home.
Probable causes:
Misaligned flash lamp
Disconnected photodiode cable
Epoch Operator’s Manual

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Biotek Epoch and is the answer not in the manual?

Biotek Epoch Specifications

General IconGeneral
TypeMicroplate Spectrophotometer
Wavelength Range200 - 999 nm
Light SourceXenon flash lamp
Power Requirements100 - 240 VAC, 50/60 Hz
Wavelength Repeatability+/- 0.2 nm
Absorbance Range0-4.0 OD
Plate Types6- to 384-well microplates
ShakingLinear, orbital
Detection ModesAbsorbance
SoftwareGen5
Bandwidth5 nm
ConnectivityUSB

Related product manuals