58 | Appendix B: Error Codes
Code Description and Probable Causes
*0400
Carrier x-axis failed positional verify.
Motor x-axis failed to reach the same position when moved a
known number of steps from the home position and back.
Probable causes:
Dirty linear bearing.
Obstruction in read chamber.
*0401
Carrier y-axis failed positional verify.
Dirty linear bearing.
Obstruction in read chamber
*0402
Order sorting filter wheel failed positional verify.
Filter wheel is loose.
Filter wheel gearing is worn or dirty.
Defective or broken optical sensor or connection.
Defective motor, motor controller PCB, or cable.
*0403
Monochromator failed positional verify.
Probable causes:
Defective measurement diode PCB or connection.
Defective mechanics or optics
Defective flash lamp.
Defective motor/power PCB or connection to it.
Flash lamp is missing flashes or is not flashing.
The optic system does not detect the saturation.
Defective monochromator (low probability).