800953 Bittium Faros Manual
34 (80)
Version 7.0.0
© Bittium Biosignals Ltd 2023
1. Run Bittium Faros Manager.
2. Select ‘Holter’ mode from Manager main
view.
3. Select ‘Custom’, or desired measurement
time (Preset recording options, 1-7 days)
from Profile Types.
4. When using “Custom settings”, select
wanted measurement parameters:
a. Patient category, Adult/Pediatric
b. RR Intervals, On/Off
c. Accelerometer, On/Off
d. ECG/EKG, On/Off
5. ECG Channel count (3 channels option is on-
ly available with Faros 360).
6. Select the electrode positions to match the
electrode type and positioning used for the
measurement.
7. Select values for acquisition parameter:
a. ECG - Sampling Frequency
b. Accelerometer - Sampling Frequen-
cy
c. Accelerometer - Dynamic Range
8. Autostart on skin contact, On/Off
9. Temperature measurement On/Off (ONLY
with Faros 360).
10. Press ‘Save’ to apply the configuration.
11. Click “Eject” to eject Faros device.
NOTE
The selection of Electrode positions affects the displayed device settings. Set the Electrode
positions picture to match the type of electrode used for measurement.
6.2 Cardiac Event Recorder configuration
Cardiac event recordings can be performed with every Faros model. The configuration for Cardiac Event
Recorder mode is set using Faros Manager.