Table 10 MiniSnap Sensitive electrode positioning
MiniSnap Sensitive 1-CH Electrode
NOTE
Bittium MiniSnap Sensitive 1-CH ECG electrode is the most suitable for 24h ECG
measurements on small adults and children.
NOTE
Optimal signal quality is obtained when the electrode is fully placed on top of the thorax
and the bottom of the electrode is clearly above the diaphragm.
7.2 Patient preparation for electrodes and leads
Step 1: Prepare the skin of the patient in areas where the electrodes are placed:
a) Shave the hair off from the areas where the electrodes are placed.
b) Clean the skin with appropriate alcohol (for example denat. 80% alcohol), electrode preparation
pads or using mild soap and water.
Step 2: Attach the electrodes to the patient’s skin:
c) Electrode placement is instructed in next chapter.
d) Check that the adhesive sticks properly.
e) Check that there is no air between the electrode gel and patient’s skin.
f) Check that there is no hair under the adhesive.
Step 3: Attach the device to the cable set and the cable set to the electrodes: