800953 Bittium Faros Manual
© Bittium Biosignals Ltd 2023
CAUTION: Use of non-authorized accessories may break the OmegaSnap electrode.
CAUTION: Conductive parts of electrodes and associated connectors for applied parts, in-
cluding the neutral electrode, should not contact other conductive parts including earth.
CAUTION: Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to any part
of the device, including cables specified by Bittium Biosignals Ltd. Otherwise, degradation of
the performance of this equipment could result.
Leakage currents on patient connections are limited under acceptable levels if there are
short circuit in the OmegaSnap electrode input/output parts.
EMC disturbances might cause interference and/or noise to measurement data.
OmegaSnap electrode is not reusable. Electrode performance degradation or contamination
may occur with a used OmegaSnap electrode.
Always follow instructions for disposable electrodes which are used for recordings.
Any incident (serious or non-serious) that has occurred in relation to the device shall be
reported to the manufacturer and the competent authority of the Member State in which
the user and/or patient is established.
The electrode is CE-marked for the conformity to Council Regulation 2017/745 regarding
medical devices.