Be careful not to rub too forcefully. Do not use alcohol-based fluids or corrosive chemicals!
Do not sink any cables in water! Do not rinse any cables or measurement devices with
cleaning fluid!
9.2.2 OmegaSnap adapter
OmegaSnap adapter can be cleaned by wiping it using a non-fluffing cloth dampened with cleaning fluid
such as mild hand soap solution, or water. A non-fluffing cloth dampened with non-alcoholic disinfection
fluid can be used for disinfection. Recommended way of cleaning: non-alcoholic cleaning and disinfection
wipes for medical devices, for example mikrozid® sensitive wipes.
Do not use cleaning fluid which includes ethers, ketones or partially halogenated or
aromatic hydrocarbons! Make sure that the USB connector is dry before use!
Be careful not to rub too forcefully. Do not use alcohol-based fluids or corrosive chemicals!
Do not rinse measurement devices with cleaning fluid!
9.3 Device battery replacement
Faros device battery is an in-built part of the device and can be changed only by Bittium Biosignals Ltd. Bat-
tery lifetime depends on device usage modes and recharging cycles. It is recommended to replace the bat-
tery after 300 cycles and in normal use scenarios this is in about 2,5 – 3 years.
When battery replacement is needed, please contact your local distributor or Bittium Biosignals Ltd.