800953 Bittium Faros Manual
76 (80)
Version 7.0.0
© Bittium Biosignals Ltd 2023
CAUTION: Bittium Faros 180L does not directly provide diagnosis as a supervising physician
is responsible for ECG data interpretation.
CAUTION: If a patient has been given defibrillation, while Faros ECG device and applied part
are connected to the patient, Faros device and applied part must be sent to manufacturer
for checking before continuing device use.
CAUTION: Bittium Faros 180L device is not suitable to use in MRI environment.
CAUTION: Bittium Faros 180L device is not intended to be used at the same time with high
frequency (HF) surgical equipment or with defibrillator.
CAUTION: Patients, who have active implantable medical device (like heart pacemaker etc.)
should consult supervising physician or doctor before use.
CAUTION: Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to any part
of the device, including cables specified by Bittium Biosignals Ltd. Otherwise, degradation of
the performance of this equipment could result.
CAUTION: Use Faros ECG device with accessories provided by Bittium Biosignals Ltd. Other
cables and accessories may negatively affect device performance.
NOTE
EMC disturbances might cause interference and/or noise to measurement data.