800953 Bittium Faros Manual
9 (80)
Version 7.0.0
© Bittium Biosignals Ltd 2023
CAUTION: Use ECG sampling rate of 500Hz with pediatric patients weighing less than 10 kg.
CAUTION: If a patient has been given defibrillation, while Faros ECG device and applied part
are connected to the patient, Faros device and applied part must be sent to manufacturer
for checking before continuing device use.
CAUTION: Faros device is not suitable to use in MRI environment.
CAUTION: Faros device is not intended to be used at the same time with high frequency
(HF) surgical equipment or with defibrillator.
CAUTION: Patients, who have active implantable medical device (like heart pacemaker etc.)
should consult supervising physician or doctor before use.
CAUTION: Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to any part
of the device, including cables specified by Bittium Biosignals Ltd. Otherwise, degradation of
the performance of this equipment could result.
CAUTION: Use Faros ECG device with accessories provided by Bittium Biosignals Ltd. Other
cables and accessories may negatively affect device performance.
NOTE
EMC disturbances might cause interference and/or noise to measurement data.
NOTE
Any incident (serious or non-serious) that has occurred in relation to the device shall be
reported to the manufacturer and the competent authority of the Member State in which
the user and/or patient is established.
NOTE
Faros device does not have any electrical stimulation capabilities.
NOTE
Accelerometer data is not analyzed within the device or differentiated between various