107
Startup Times
List the number of startup times.
Software version
List the software version.
Hardware Version
List the hardware version.
Product type
List the product model.
Serial No.
List the product serial number.
Language
Menu items can be displayed in 12 different languages, selectable by the
user.
Press the “UTILITY” button and then select “Language”.
Figure 3.59 Language Selection Screen
Self Calibration
Self Calibration procedure can optimize the signal path for optimal
measurement results. You can run this procedure at any time. If the
operating temperature changes by more than 5° C or the instrument runs
more than thirty minutes, running self calibration is recommended.
When you run the self Calibration, you must disconnect all the probes and
leads, and make sure the scope is not in Single-Shot mode. Then press the
“Utility” button to choose the “Do self cal” to show the self Calibration