SAM 950 Berkeley Nucleonics Corporation 24 of 62
Figure 4.10. Background & Calibration screen information
Background measurement is important for accurate nuclide identification. Ensure that no radiation
sources are in the vicinity of the SAM 950 during the background measurement.
In calibration mode, place a Cs-137 source front the SAM 950. When the slected number of counts
have been acquired, the screen will display “Success”. Remove the Cs-137 source when calibration is
completed. The x-axis unit change from channel number to energy in keV.