56 en | NurseCall Main Unit
F.01U.252.699 | V1.1 | 2012.01 User Manual Bosch Security Systems
The x digit can be programmed. Section Specific parameter for POCSAG, page 30.
Change day / night
If the change day / night is activated, the NurseCall system transfers all alarms during night to
the group 24. During the day, all groups 00 - 24 can be used. When switching from day to
night or vice versa, the message "Day-Night" is sent to the activated pagers.
Priority alarms
The assistance and fire alarms are priority calls sent to all activated pagers.
Call repetition
If alarms or messages are not acknowledged after approximately 7 minutes, call repetitions
are also sent to the group 23.
A.5.3 DeTeWe protocol
Characteristics
Data rate: 9600 Bauds
Transmission: asynchronous
11 bit-structure (1 start bit, 8 data bits with odd parity, 1 stop bit)
half-duplex mode.
Only alarms or messages arriving to the alarm buffer are transmitted, except "Relay Off". The
transmission is fully alpha-numeric. Acknowledgement messages are also transmitted.
Alarms are sent to the DeTeWe user number attributed to the transmitter.
The default group is 00.
16 001x128
17 001x136
18 001x144
19 001x152
20 001x160
21 001x168
22 001x176
23 001x184
24 001x192
Group User number
Group User number
00 x00
01 x01
02 x02
03 x03
04 x04
05 x05
06 x06
07 x07
08 x08