10 Technical specifications
108 Operating Manual ARC 350 900-351_IFU_V2.0.0_20221-S0-20150520-EN
Bipolar Cutting Standard
Measurement at ohmic resistances
Diagram of power output P [W] as a
function of the load resistance R [] for
the setting "Bipolar Cutting Standard"
= 100 W
Measurement at ohmic resistances
Diagram of power output P [W] as a
function of the load resistance R [] for
the setting "Bipolar Cutting Standard"
= 200 W
Diagram of power output P [W] as a
function of the setting "Bipolar Cutting
Standard"
Rated load resistance = 75
HF output voltage U [Vp] for the setting
"Bipolar Cutting Standard" (idle mode)
= 400 Vp