10 Technical specifications
900-351_IFU_V2.0.0_20221-S0-20150520-EN Operating Manual ARC 350 75
Neutral electrode monitoring
EASY (Electrode Application System)
Display indication of one-piece or split or Baby
electrode in the main and neutral electrode menu
Contact resistance between individual sections of
split neutral electrodes shown on display with
column for control
Lead resistance shown on the display when a one-
piece neutral electrode is used
Maximum allowable resistance between the
sections of a split electrode
300 Ω
Warning signal for hazardous conditions beneath
the neutral electrode
Visual, acoustic
Warning messages on display:
Text messages with further
information
Safety features
ISSys (Integrated Safety System)
Continuous monitoring of HF leakage current with
fault indication
Text messages with further
information
Dosage monitoring with fault indication on the
display
Continuous self-test
Continuous status indication on the display
Operating errors shown on the display
Text messages with further
information
System faults shown on the display
Text messages with further
information
Documentation
Data acquisition and storage in the device
System information with date
Fault states
Operating errors
Data retrieval via the display
Text messages with further
information
Communication
External interface for communication with ARC
PLUS (light wire cables)
USB interface for software updates
External PC interface, UART, using BOWA
software
Service support using BOWA software

Table of Contents