EasyManua.ls Logo

Bowa ARC 350 - Page 76

Bowa ARC 350
124 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 Technical specifications
76 Operating Manual ARC 350 900-351_IFU_V2.0.0_20221-S0-20150520-EN
Service support
Network port for service support
Service support by service programs integrated in
the device
Service support via ISSys
Cooling
Convection
Temperature-controlled fan
Duty factor
Intermittent
10 s / 30 s (on/off)
Characteristics
Max. CUT Power (at 500 Ω)
350 W
Max. COAG Power (at 25 Ω)
350 W
Monopolar sockets
2 x international / Erbe
Bipolar sockets
2 x international / Erbe
Connection for footswitch
2 x
AUTOSTART
Bipolar finger switch
Plug’n Cut COMFORT instrument identification
Optionen
Argon / GastroCut
REF 900-391
Bipolar Resection
REF 900-395
LIGATION
REF 900-396
Umweltbedingungen
Betrieb
Transport und
Lagerung
Temperature
+10°C to +40°C
-20°C to +50°C
Relative humidity
30 to 75%, non-
condensing
0 to 90%, non-
condensing
Atmospheric pressure
700 to 1060 hPa
500 to 1060 hPa
Operating altitude (max)
4000 m a. sea
level