10 Technical specifications
94 Operating Manual ARC 350 900-351_IFU_V2.0.0_20221-S0-20150520-EN
Monopolar Coagulation Moderate
Measurement at ohmic resistances
Diagram of power output P [W] as a
function of the load resistance R [] for
the setting "Monopolar Coagulation
Moderate"
= 60 W
Measurement at ohmic resistances
Diagram of power output P [W] as a
function of the load resistance R [] for
the setting "Monopolar Coagulation
Moderate"
= 120 W
Diagram of power output P [W] as a
function of the setting "Monopolar
Coagulation Moderate"
Rated load resistance= 75
Effect
U (Vp)
1
250
2
250
3
250
Table of HF output voltage U [Vp] as a
function of the setting "Monopolar
Coagulation Moderate"
(idle mode)