Disassembly / Assembly
3 - 38 Infusomat® Space 6.0
EN
Mounting Connector Seal P2 (self-adhesive)
Designation Order No.
Connector seal P2 SP
3477 3102
1. Clean the adapter (Fig
.: 3 - 39 / Item 1) of the connector seal
P2 all round with isopropanol and let it air.
2. Preposition the connector seal P2 (Fig.: 3 - 40 / Item 1) (incl.
inner cut-out).
Fig.: 3 - 39
Legend of fig. 3 - 39:
Item Designation
1 Adapter
Fig.: 3 - 40
Legend of fig. 3 - 40:
Item Designation
1 Connector seal P2