PULSE Multi-analyzer System Type 3560-B/C/D/E Installation and IDAe Hardware120
Specifications – Charge Input Channels, Dyn-X
(all specifications for transducer capacitance = 1 nF)
Dyn-X
3035 (TNC)
a
Frequency Range 0.1 Hz to 25.6 kHz
A/D Conversion 2 × 24-bit
Data Transfer 24-bit
Input Range 2 ranges: 10 nC
peak
/1 nC
peak
Input Signal Ground Coupling Floating or single-ended (grounded to chassis)
Absolute Maximum Input ±300 nC
peak
without damage
High-pass Filters
b
(see Fig.11.2)
DC (f
L
= 0) Not available
10% –10% 3 dB Slope
0.1 Hz high-pass filter 0.1 Hz 0.065Hz 40 dB/dec.
1 Hz high-pass filter 1.0 Hz 0.65 Hz 40 dB/dec.
10 Hz high-pass filter 10.0 Hz 7.88Hz 60 dB/dec.
30 Hz high-pass filter 30.0 Hz 23.63 Hz 40 dB/dec.
Low-pass Filters
(see Fig.11.3)
10% –10% 3 dB Slope
100 Hz low-pass filter 100 Hz 143Hz 40 dB/dec.
1 kHz low-pass filter 1.0kHz 1.43 kHz 40 dB/dec.
3 kHz low-pass filter 3.0 kHz 4.31kHz 40 dB/dec.
10 kHz low-pass filter 10.0 kHz 14.3 kHz 40 dB/dec.
30 kHz low-pass filter 30.0 kHz 43.1 kHz 40 dB/dec.
low-pass filter bypassed See Anti-aliasing Filter
Absolute Amplitude Precision, 1kHz, 1 V
input
±0.05dB, typ. ±0.01dB
Attenuator Linearity @ 1 kHz ±0.05 dB, typ. ±0.005 dB
Amplitude
Linearity
(linearity in one
range)
0 to 60 dB below full scale ±0.05 dB, typ. ±0.01 dB
60 to 80 dB below full scale ±0.05 dB, typ. ±0.01 dB
80 to 100 dB below full scale ±0.2 dB, typ. ±0.02 dB
100 to 120dB below full scale typ. ±0.02 dB
120 to 140dB below full scale typ. ±0.02 dB
140 to 160dB below full scale typ. ±1 dB
Overall Frequency Response
re 1 kHz, lower limit f
L
to upper limit f
U
Min. 0.1 Hz to max. 25.6 kHz: ±0.1 dB, –10% at f
L
Noise: Measured lin. 10 Hz to
25.6 kHz: fC
rms
(input terminated by 1nF)
(Values in parentheses are
specified in
aC
rms
/Hz [a = 10
–-18
])
Input Range Signal Level Guaranteed Typical
1nC <316pC
peak
<5 (<32) <3 (<19)
1nC >316pC
peak
<15 (<94) <7.5 (<47)
10 nC <3160 pC
peak
<15 (<94) <9 (<57)
10 nC >3160 pC
peak
<65 (<407) <50 (<313)
Spurious-free Dynamic
Range
(dB) re full scale input
(input terminated by1nF)
Input Range Typical
1nC
peak
140 dB
10 nC
peak
150 dB
DC Offset re full scale Not applicable

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BRUEL & KJAER PULSE 3560-C and is the answer not in the manual?

BRUEL & KJAER PULSE 3560-C Specifications

General IconGeneral
BrandBRUEL & KJAER
ModelPULSE 3560-C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals