Scheduled maintenance
CryoProbe User Manual BRUKER 69 (107)
NOTE: If more than 10 000 operating hours elapsed on the CryoCooling Unit’s
counter since the last regular service, some parts may have worn out seriously.
An in-depth inspection and service will be required that takes more than two days.
IMPORTANT: If no scheduled maintenance is carried out, individual components
will eventually fail due to aging or wear-out, and they may even cause substantial
damage to other parts. However, under virtually no circumstances a hazard for
the user would be present.
Exchange of Adsorber in He Compressor
Needed every 20 000 h. Since a specialist must visit each 10 000 h anyhow for
the Coldhead exchange, Adsorber and Coldhead exchange can be combined.