Water chiller (optional)
CryoProbe User Manual BRUKER 23 (107)
Technical data
Data for TXI configuration (cover and 5 HPPR modules), an X-BB module is in-
cluded by default:
Dimensions (w × d × h in mm) 260 × 350 × 420
Weight 25 kg
Water chiller (optional) 3.4
0
The He Compressor generates about 7.5 kW of heat. If water-cooling has been
chosen but no cooling water is available, a water chiller should be used. Fresh
water is not recommended because of cost, purity, and environment protection.
Technical data
The following example is valid for the water chiller available from B
RUKER in Eu-
rope. It is of the ‘split type’, i.e. it consists of two units: a main unit indoors with
pumps for a primary and a secondary cycle linked by a heat exchanger, and a ra-
diator outdoors that disposes the heat from the secondary cycle to the atmo-
sphere. Local suppliers in other parts of the world are likely to offer similar
devices.
Dimensions (w × d × h in mm)
main cabinet (indoors) 690 × 920 × 1490
radiator (outdoors) 1895 × 470 × 832
Weight
main cabinet (indoors) (120 kg empty) 250 kg
radiator (outdoors) (94 kg empty) 120 kg
Figure 3.5. Cryo-compatible preamplifier assembly ‘HPPR CRP’
‘RF in’ from
CryoPreamps
1H CRP module
2H CRP module
13C CRP module
15N CRP module
conventional X-BB
transmit & receive,
decoupling channel
transmit only,
observe channel
CryoProbe power
supply on back