EN-42 T104 Series Indicators
8.2 Appendix B
MT-SICS Commands
Command Function
LEVEL 0
@ Reset the scale
I0 Inquiry of all available SICS commands
I1 Inquiry of SICS level and SICS versions
I2 Inquiry of scale data
I3 Inquiry of scale software version
I4 Inquiry of serial number
S Send stable weight value
SI Send weight value immediately
SIR Send weight value repeatedly
Z Zero the scale
ZI Zero immediately
LEVEL 1
D Write text into display
DW Weight display
SR Send and repeat stable weight value
T Tare
TA Tare value
TAC Clear tare
TI Tare immediately
Command Function
LEVEL 2
C2 Calibrate with the external calibration weight
C3 Calibrate with the internal calibration weight
I10 Inquire or set scale ID
I11 Inquire of scale type
P100 Print out on the printer
P101 Print out stable weight value
P102 Print out current weight value immediately
SIRU Send weight value in the current unit immediately and repeat
SIU Send weight value in the current unit immediately
SNR Send stable weight value and repeat after every weight change
SNRU Send stable weight value in the current unit and repeat after every weight
change
SRU Send weight value in the current unit and repeat
ST After pressing the Transfer key, send the stable weight value
SU Send stable weight value in the current unit
LEVEL 3
M01 Weighing mode
M02 Stability setting
M03 Autozero function
M19 Send calibration weight
M21 Inquire/set weight unit
PRN Print out at every printer interface
RST Restart
SFIR Send weight value immediately and repeat quickly
SIH Send weight value immediately in high resolution
SWU Switch weight unit
SX Send stable data record
SXI Send data record immediately
SXIR Send data record immediately and repeat
U Switch weight unit